The Global FIA miniSIA-1 represents a new generation of instrumentation for the proven Sequential Injection technique. It combines sampling frequency and speed of response of FIA with the robustness and reagent economy of SIA. The key to this new capability is the marrying of the Global FIA milliGAT pump with flow programming. The concept was first described by Ruzicka in the FIA Tutorial.
The system is equipped with a milliGAT pump coupled to a heating coil and Chem-on-Valve™ (COV) monolithic manifold. An extra measure of modularity is built into the COV by making the detector flow cell a bolt-on piece. This means that the same manifold can be configured with a short or long path length absorbance cell, fluorescence cell, chemiluminescence cell, and one of a host of different electrochemical detectors. External instruments such as Atomic Absorption, Inductively Coupled Plasma and Electrospray MS are compatible with miniSIA-1 since the milliGAT pump provides a wide range (6nL/sec to 170 µL/sec) of highly reproducible flow rates over long periods of time.
For spectrophotometry the standard 10 mm uv/vis absorbance cell uses fiber optic (f.o.) cables to pipe light to and from the flow cell. A white LED and a WXe lamp are provided as light sources. The spectrometer is an Ocean Optics USB-4000.
The sample port has an inlet and outlet which meet on the face of the COV rotor. Thus the sample can be primed to the valve without drawing sample into the rest of the manifold where it can be a source of carryover. The sample port can also be primed with the next sample while the present measurement is underway, either manually or by a miniaturized metering pump.
Power supply voltage |
100 to 240 VAC, 50/60 Hz |
Power consumption |
100-240 VAC, 10A (max) |
Sample size |
10-50 µL |
Heater temperature range |
Ambient to 120 °C |
Light source |
White LED and W-Xe lamp |
Spectrometer wavelength range |
350 to 1050 nm |
Software |
FloZF |
Available Chemistries |
PO4, NOx,NH3,SO4,SiO4,Fe, Mn, Al, Zn, Enzymatic assays |
Sequence time |
30-360 seconds |
Precision mid calibration range |
1% |
Linearity of calibration curves |
r2:0.99 |
Weight |
5 kg |
Dimensions |
24 x 23 x 26 cm (W x D x H) |
Ambient operating temperature |
4 to 60 °C |
Ambient operating humidity |
25% to 85% non-condensing |
Storage temperature |
-25 to 65 °C non-condensing and non-icing |
Communications |
USB 2.0 and Wi-Fi 802.11ac |
COV-FL1: Fluorescence flow cell with LED excitation
COV-CL: Chemiluminescence flow cell
PC-P25232: Photon counter
SI-CHROM: SI-Chrom Kit
SIA-KIT: SIA Fittings kit
COV-LPFC10-H: Heated long path length photometric flow cell, 10 cm
Brochure: miniSIA-1 brochure.pdf611.74 KB
Manual: man_miniSIA-2_-.pdf1.91 MB
Manual FloZF: man_FloZF_5.2.pdf3.16 MB
Manual FloZF Getting Started: man_FloZF_Getting_Started_-.pdf1.16 MB
Miniaturized SIA Instruments and Components
Aluminum Assay by Fluorescence
Zn in Sea Water by Fluorescence Detection Using SI-LOV Platform
In the early days of FIA, the most fertile repository of manifold components was the Chromatographer's laboratory. Since those early days, a whole suite of custom components have been developed for those practicing flow-based analysis. These components are now available commercially as stocked items. Global FIA provides a single source for many of the components needed to equip and maintain a flow-based analysis laboratory.
The range of components offered includes:
In recent years we have added a number of sampling components as well. These include:
All of these devices are available in the Global FIA store.
684 6th Ave, Fox Island WA, 98333
+1(253) 549 2223 click to call
info@globalfia.com
Monday - Friday
9am - 5pm (PST)